Brake Pedal Position Sensor Calibration
Calibration Criteria
Note: Do not apply the brake pedal during the brake pedal position
(BPP) sensor calibration procedure. Any movement of the brake pedal
during this procedure will cause the calibration procedure to fail. If this
occurs, the sensor calibration must be repeated.
Brake pedal ...
Installation Procedure
Cut the front hinge pillar body in corresponding locations to fit the
remaining original panel. The sectioning joint should be trimmed
to allow a gap of one-and-one-half-times the metal thickness at the
sectioning joint.
Create a 50 mm (2 in) backing plate from the unused portion of ...
Air Hose Test
Caution: The air hose test should only be used on fully cured
urethane adhesive. Otherwise, damage to the urethane adhesive bead
could result in additional leaks.
Using a liquid detergent, diluted with water in a spray bottle, spray
the window at the edges. Begin at the bottom and grad ...